ClearanceBrief
Recall Z-0568-06 ยท LFR Glucose Dehydrogenase, Glucose

Roche Diagnostics Corp.

Accu-Chek Advantage II blood glucose test strips; Reference numbers 12030659001, 12030667001, 03137872001, 03137899001, 03000281001, 12030535001, 12030543001, 03033449001, 03033465001 and 12030543047 (All sold outside of the United States).

DateFeb 28, 2006 (initiated Feb 10, 2006)
StatusTerminated
ReasonThere may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
Root causeOther
ActionNotifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.
DistributionNationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan and Thailan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982089ACCU-CHEK ADVANTAGE TEST STRIPS AND ACCU-CHEK ADVANTAGE H TEST STRIPS
Boehringer Mannheim Corp., Sep 3, 1998