Recall Z-1266-06 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 4 inch diameter Roller Pump; Model 801040.

DateJul 27, 2006 (initiated Sep 1, 2005)
StatusTerminated
ReasonThe pump may stop during the autodose delivery without completion of the dose delivery to the patient.
Root causeOther
ActionService technicians were notified on 9/30/05 to upgrade the software on these units by March 31, 2006 as follows: pump software to version 1.20, power manager software to version 1.20 and monitor software to version 2.2.1. U.S. software has all been upgraded and international upgrades are ongoing as of 7/7/06.
Quantity1898 combined
DistributionNationwide, and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand, Turkey and United Arab Emirates.