Recall Z-1207-06 ยท INT Plinth

SAMMONS PRESTON ROLYAN

ValueLine Mat Platform Model C2230A,B,C; C922-30A, B,and C; C6904, C6903, C6902, C6901. All tables have the standard platform and the model numbers represent the additional shelves or drawers and the arrangement.

DateJul 12, 2006 (initiated Jan 25, 2006)
StatusTerminated
ReasonThere have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
Root causeOther
ActionLetters were sent to customers certified mail on 1/25-26/2006 instructing them to take the tables out of service until it could be inspected by a sales representative. Sales represenatives are following up with customers in person to inspect tables.
Quantity561 devices total
DistributionNationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.