Recall Z-1286-06 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801763 and 801763 (Japan unit).

DateJul 27, 2006 (initiated Oct 19, 2005)
StatusTerminated
ReasonWhile on battery power, the system may flash a low battery warning even though the battery is adequately charged.
Root causeOther
ActionCustomers were notified via letter dated October 19, 2005 that the Power Manager Software would be upgraded within the next six months. As of 7/7/06 all units within the U.S. have been completed and international units are still in the process of being upgraded.
Quantity329
DistributionNationwide and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand, Turkey and United Arab Emirates.