Recall Z-0528-06 ยท CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone

Diagnostic Products Corp.

IMMULITE 2000 Total Testosterone Kit

DateFeb 16, 2006 (initiated Jan 3, 2006)
StatusTerminated
ReasonThe reagent for this kit lot (L2KTW 2,6 126) was manufactured with a raw material that can carryover and interfere with DPC Kit IMMULITE 2000 Estradiol (L2KE2) causing a false elevation in the observed estradiol concentration.
Root causeOther
ActionFor domestic customers - the recall is extended to the domestic end user-level. Domestic customers were notified via telephone using a Phone Script (attached in Attachment 5), starting on January 5,2005. Each customer receives a fax-back form and a Technical Bulletin directly after this phone contact. For foreign consignees - the recall is extended to foreign distributors. It is the distributor''s responsibility to extend the recall action to their direct customers. Foreign consignees were notified via e-mail and/or FAX,starting on January 3, 2005.
Quantity710
DistributionNationwide and Worldwide in Pakistan, Italy, Cyprus, Turkey, Switzerland, Chile, South Africa, Greece, Mexica, Canada, Thialand, Germany, India, United Arab Emirates, Hungary, Israel, Portugal, Taiwan, Austria, Saudi Arabia, Malaysia, Hong Kong, India, Colombia, Argentina, Australia, Costa Rica, Cze
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032881IMMULITE/IMMULITE 1000 TOTAL TESTOSTERONE AND IMMULITE 2000 TOTAL TESTOSTERONE
Diagnostic Products Corp., Oct 10, 2003