Recall Z-0565-06 ยท DPS Electrocardiograph

Inovise Medical, Inc.

AUDICOR 200 D Inovise Part Number 30047 (console part number 20158).

DateMar 3, 2006 (initiated Jan 12, 2003)
StatusTerminated
ReasonUse of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power.
Root causeOther
ActionOn January 12, 2006, 2 of 3 customers were notified by telephone, with follow-up email. All three customers were visited by the firm on January 16-17, 2006. Customers were instucted to not use the device in-line with a defibrillator/monitor.
Quantity28 units
DistributionProduct distributed to three EMS/Fire Departments located in Florida and Washington
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043074LIBERTY SYSTEM