Recall Z-0515-06 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Babylog 8000 plus with software version 5.n. Neonatal Respirator-Continuous Ventilator

DateFeb 14, 2006 (initiated Dec 28, 2005)
StatusTerminated
ReasonFailure to generate an alarm when an unintended disconnection occurrs
Root causeOther
ActionRecalling firm has issued a medical device recall letter to consignees to the attention of the Director, Neonatal Intensive Care Unit via Certified Mail to alert them of additional monitoring requirements if a PEEP setting of less than 2.5 mbar is used
Quantity2261 units
DistributionProduct was shipped to medical facilities nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974176BABYLOG 8000 PLUS