Recall Z-0563-06 ยท JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use

Cygnus LLC

Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

DateFeb 28, 2006 (initiated Feb 4, 2005)
StatusTerminated
ReasonThe units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
Root causeOther
ActionA recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
Quantity20 units