Recall Z-0514-06 ยท JTC Device, Microtiter Diluting/Dispensing
Hamilton Co
Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036
| Date | Feb 14, 2006 (initiated Dec 13, 2005) |
| Status | Terminated |
| Reason | Heat sinking on power boards may not operate as intended and the board may cease to operate. |
| Root cause | Other |
| Action | On 12/15/05, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned. |
| Quantity | 1071 units |
| Distribution | Product was released for distribution to 132 consignees in US, and 17 International consignees |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K871945 | MICROLAB AT Hamilton Co., May 29, 1987 |