Recall Z-0639-06 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Baxter Minivolume Extension Set, product code 1C8400; An Rx sterile, nonpyrogenic fluid pathway, 13.0'', 0.4 mL Vol., with male Luer Lock adapter; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

DateMar 24, 2006 (initiated Dec 15, 2005)
StatusTerminated
ReasonThe extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
Root causeOther
ActionBaxter sent Urgent Product Recall letters dated 8/3/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the microbore female luers that may crack during use, resulting in the possible under delivery or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter extended the recall to include
Quantity13,824 sets
DistributionNationwide, including Puerto Rico and American Samoa, and internationally to Canada, Australia, Japan, South Korea, Saudi Arabia, India, Mexico, Venezuela, Malaysia, East Timor and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K811078RADIATION STERILIZED ADMIN. SET
Travenol Laboratories, S.A., Jul 2, 1981