Recall Z-0385-06 ยท JZJ Prealbumin, Antigen, Antiserum, Control

Abbott Laboratories Diagnostic Div

Clinical Chemistry Prealbumin, List Number 1E02-20

DateJan 11, 2006 (initiated Nov 10, 2005)
StatusTerminated
ReasonThese specific protein assays are not meeting the 57-day on board stability claim
Root causeOther
ActionA Device Correction letter identifying the changes being made to the frequency of QC monitoring for these assays, will be distributed to all AEROSET and ARCHITECT c8000 customers which utilize the IgM, Prealbumin and Apolipoprotein Al Specifc Protein assays on November 10,2005 via direct mailing. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check..
Quantity734 (568 in US)
DistributionNationwide, plus Australia, Canada, Columbia, Germany, Guatemala, Hong Kong, mexico, & New Zealand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.