ClearanceBrief
Recall Z-0449-06 ยท FTL Mesh, Surgical, Polymeric

Ethicon, Inc.

PROCEED Surgical Mesh is a sterile, thin, flexible laminate mesh designed for the repair of hernias and other fascial deficiencies. The mesh product is comprised of an oxidized regenerated cellulose (ORC) fabric, and PROLENE Soft Mesh, a nonabsorbable polypropylene mesh, which is encapsulated by a polydioxanone polymer. The polypropylene mesh side of the product allows for tissue ingrowth while

DateJan 26, 2006 (initiated Jan 4, 2006)
StatusTerminated
ReasonLots of Proceed Surgical Mesh may delaminate from the polypropylene mesh during certain hernia repairs. Exposure of the polypropylene layer to the bowel prior to reperitonealization could increase the risk of adhesions and bowel fistulization.
Root causeOther
ActionThe recall was initiated on 12/21/2005. The recall to the customers were initiate via letter with return receipt on 1/4/2006.
Quantity18,270 devices
DistributionThere is a total of 1,070 consignees that have received recall product divided between hospitals and distributors nationwide. The product has been distributed internationally to the following countries: Argentina, Australia, Austra, Brazil, Canada, Chile Colombia, Czechoslovakian Republic, France,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031925PROCEED TRILAMINATE SURGICAL MESH
ETHICON, Inc., Sep 17, 2003