ClearanceBrief
Recall Z-0381-06 ยท LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers

Roche Diagnostics Corp.

Roche Elecsys total PSA, catalog number 11731262322.

DateJan 11, 2006 (initiated Dec 7, 2005)
StatusTerminated
ReasonThere is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
Root causeOther
ActionConsignees were notified of the incompatibility via reagent bulletin dated 12/7/05.
Quantity25,000 kits
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.