Recall Z-0398-06 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHITECT c8000 System Software, Version 2.11, Product List/Model #5F48-12, distributed by Abbot Laboratories, Irving, TX.

DateJan 18, 2006 (initiated Dec 8, 2005)
StatusTerminated
ReasonSystem software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL.
Root causeOther
ActionFirm initiated recall to all consignees via letter dated 12/08/05.
Quantity1206 units.
DistributionWorldwide. Product distributed nationwide to direct accounts (including gov''t accounts) in: AL, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MN, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WY. Foreign distribution to the following countries: Argentin