Recall Z-0470-06 ยท GHM Test, Sickle Cell

Chembio Diagnostic System Inc

Sickle-STAT Test Kit, Catalog # SC901, a qualitative test kit for the determination of the presence of hemoglobin S in human blood --- FOR IN VITRO DIAGNOSTIC USE --- FOR PROFESSIONAL USE ONLY --- Contents: 48 pre-filled Sickle-STAT reaction tubes containing sodium hydrosulfite. 1 bottle phosphate buffer containing surfactant and stabilizer, and 1 Instruction leaflet. --- manufactured by: CHEMBIO

DateFeb 4, 2006 (initiated Dec 16, 2005)
StatusTerminated
ReasonFirm received one report of a false negative with the positive control. Internal testing by the firm has shown that the kit lot may result in invalid or false negative interpretations.
Root causeOther
ActionChembio phoned the consignees and followed up by faxing the recall letters (dated 12/15/05) & response forms on 12/22/05.
QuantityLot SC050305: 347 test kits, Lot SC062805: 410 test kits
DistributionMedical facilities and a few distributors located nationwide, the Dominican Republic, and West Indies (St. Vincent & Grenadine)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K841272SICKLE-STAT HEMOGLOBIN S SCREEN KIT
Clinical Technology Corp., Apr 13, 1984