Recall Z-0397-06 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Plus Orthopedics USA

VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument)

DateJan 18, 2006 (initiated Dec 2, 2005)
StatusTerminated
ReasonThe adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
Root causeOther
ActionThe adjustment know tower on one device was found t have broken off from the body of the instrument. This did not happend during a surgery, and is belived tohave been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
Quantity14
DistributionArizona, California, Colorado, Idaho, New York, Oklahoma, Texas, Utah, Wisconsin