Recall Z-0392-06 ยท FFT Electrode, Ph, Stomach

Medtronic Gastroenterology / Urology

Medtronic Slimline Single-Use Internal Reference pH Catheter for Gastroesophageal pH measurements. Models 9012P3101, 9012P3111, 9012P3121, 9012P3131 and 9012P3141. Distributed by / Medtronic Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA

DateJan 13, 2006 (initiated Dec 9, 2005)
StatusTerminated
ReasonCertain lots have a manufacturing defect that allows the water pushed through the catheter to travel back into the connector of the Digitrapper pH recorder. Medtronic has determined that patient safety is not compromised, but catheter function may be affected and damage may result to the Digitrapper pH recorder.
Root causeOther
ActionLetters will be sent to distributors, end-users and Medtronic Field Personnel. The end-user will be asked to return via fax a confirmation/customer receipt. Letters identify the affected catalog number and lot numbers. Customers are asked to return affected product and will receive replacement product.
Quantity140 catheters
DistributionCA, CO, FL, MA, OH, PA. OUS to include Finland, Italy and Portugal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.