Recall Z-0327-06 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)

Ortho-Clinical Diagnostics

Vitros Immunodiagnostic HBsAg CONFIRMATORY KIT * REF 680 1324. Firm on label: Ortho-Clinical Diagnostics, 62 The Broadway, Amersham, Bucks, UK. UPC *+H221680132412* . Each kit contains: 1 x kit of reagents, and protocol card for HBsAg Confirmatory Kit. According to the firm (but not on label), there are 60 neutralization reactions per confirmatory kit.

DateJan 6, 2006 (initiated Dec 15, 2005)
StatusTerminated
ReasonIncreased background signal of the sample diluent in the affected Confirmatory Kits could cause some results to be classified as 'Not Confirmed'.
Root causeOther
ActionLetters sent overnight mail on 12/15/2005. Customers instructed to discontinue use and discard remaining inventory. In a separate Q&A sheet, Ortho recommends that previously reported results be reviewed.
Quantity436 kits
DistributionNationwide distribution to end users (medical facilities, testing labs and public health agencies). One J&J foreign affiliate in Puerto Rico with sub-distribution in the Caribbean.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.