Recall Z-0327-06 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)
Ortho-Clinical Diagnostics
Vitros Immunodiagnostic HBsAg CONFIRMATORY KIT * REF 680 1324. Firm on label: Ortho-Clinical Diagnostics, 62 The Broadway, Amersham, Bucks, UK. UPC *+H221680132412* . Each kit contains: 1 x kit of reagents, and protocol card for HBsAg Confirmatory Kit. According to the firm (but not on label), there are 60 neutralization reactions per confirmatory kit.
| Date | Jan 6, 2006 (initiated Dec 15, 2005) |
| Status | Terminated |
| Reason | Increased background signal of the sample diluent in the affected Confirmatory Kits could cause some results to be classified as 'Not Confirmed'. |
| Root cause | Other |
| Action | Letters sent overnight mail on 12/15/2005. Customers instructed to discontinue use and discard remaining inventory. In a separate Q&A sheet, Ortho recommends that previously reported results be reviewed. |
| Quantity | 436 kits |
| Distribution | Nationwide distribution to end users (medical facilities, testing labs and public health agencies). One J&J foreign affiliate in Puerto Rico with sub-distribution in the Caribbean. |
| FDA record | Open on fda.gov |