Recall Z-0905-06 ยท GZB Stimulator, Spinal-Cord, Implanted (Pain Relief)

Medtronic Inc, Neurological & Spinal Division

Medtronic Model 37711 Restore Neurostimulator. The Restore Neurostimulator is an implanted, rechargeable neurostimulator which provides pain management therapy.

DateMay 25, 2006 (initiated Nov 22, 2005)
StatusTerminated
ReasonA limited number of Model 37711 Restore Neurostimulators have an incorrect internal memory parameter that causes the device to indicate it is fully recharged when it may only be 80% to 90% recharged. This anomaly does not affect patient safety. This anomoly can be corrected by reprogramming the Restore Neurostimulator using specific software.
Root causeOther
ActionA 'Customer Communication' letter was sent to Medtronic field personnel to hand-carry into the affected accounts; customer contact began on 11/22/2005. The letter explains the patients implanted with these devices likely will only be able to recharge their neurostimulator to about 80-90% of a full charge. This anomaly does not affect device safety or effectiveness but Medtronic will re-program them upon request for customer satisfaction. The letter will be followed by scheduled on-site visit to reset the internal memory parameter of the neurostimulator.
Quantity101 devices affected
DistributionNationwide (AZ, CA, FL, ID, IL, IN, KS, KY, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, and WI).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.