Recall Z-0335-06 ยท IYB Tube Mount, X-Ray, Diagnostic

Siemens Medical Solutions USA, Inc

3D TOP Ceiling Stand. X-Ray System model number 8773673

DateDec 31, 2005 (initiated Nov 30, 2005)
StatusTerminated
Reasonscrews on the rollers may become loose
Root causeOther
ActionThe recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.
Quantity460 units
DistributionThe products were shipped to hospitals and medical facilities nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K802200VERTIX 3D-U