Recall Z-0904-06 ยท FSA Lift, Patient, Non-Ac-Powered
Joerns Healthcare Inc.
Hoyer Hydraulic Patient Lifters model C-CB-L2, C-HLA, C-HLA-2, HML-400, C-HLA-2T and C-HLA-1. Distributed by Sunrise Medical, Inc. 7477 East Dry Creek Parkway, Longmont, CO 80503 USA.
| Date | May 25, 2006 (initiated Jan 6, 2006) |
| Status | Terminated |
| Reason | A field correction is being conducted on select models of Hoyer Patient Lifts because of hydraulic jack problems that have resulted in patient injury. The possibility exists that the mounting stud located in the bottom of the hydraulic jack could become loose allowing the jack to disengage from the mast of the lift. |
| Root cause | Other |
| Action | A Technical Service Bulletin was distributed along with an Urgent Medical Device Correction letter providing details for checking the mounting stud. UPDATE: Sunrise Medical sent a Second Notice Urgent Medical Device Correction letter, in follow-up to the January 2006 notice, to Dealer/Distributors starting 04/26/06. This Notice updates customers as to what Sunrise''s field correction will be for the Hoyer Model Hydraulic Patient Lifts. The Notice states that use of the affected devices should be discontinued until the lift has been inspected and repaired. Dealers/Distributors that have d |
| Quantity | 32,109 |
| Distribution | Nationwide, Canada, Asia, South America |
| FDA record | Open on fda.gov |