ClearanceBrief
Recall Z-0463-06 ยท MAE Occluder, Patent Ductus, Arteriosus

AGA Medical Corporation

AMPLATZER Duct Occluder, Order No. 9-PDA-005, Device Size 8/6mm, Length 7 mm. Sterile EO. Single Use Only. Manufactured by AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. The AMPLATZER Duct Occluder is a self-expandable device made from a Nitinol wire mesh. A retention skirt on the aortic side provides secure positioning in the ampulla of the ductus. As the occ

DateJan 31, 2006 (initiated Sep 6, 2005)
StatusTerminated
ReasonThe pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
Root causeOther
ActionUser facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
Quantity18,725 total
DistributionOUS to the following countries: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, India, Indonesia, Isreal, Italy, Jordan, Korea, Kuwait, Malaysia, Mexico, Morocco, Netherl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.