Recall Z-0532-06 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Toshiba American Med Sys Inc

Toshiba Digital Radiography System Model DFP8000D

DateFeb 23, 2006 (initiated Nov 15, 2005)
StatusTerminated
ReasonIn certain conditions the FOV of the lateral plane does not coincide with the X-ray exposed area, and consequently a part of the X-ray field may not be displayed in the magnification mode.
Root causeOther
ActionAn Advisory Letter will be sent to customers via certified mail. Corrected software will be provided at no charge. Customers can contact their local Toshiba Service person to schedule the software update.
Quantity3
DistributionCO, MI, & TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013608DIGITAL RADIOGRAPHY SYSTEM, MODEL DFP-8000D
Toshiba America Medical Systems, In.C, Nov 15, 2001