Recall Z-0369-06 ยท NCX System, Hypothermia, Intravenous, Cooling

Alsius Corporation

Alsius CoolGard Temperature Regulation System

DateJan 10, 2006 (initiated Nov 9, 2005)
StatusTerminated
ReasonAlsius has found that fluid spills or ingress may result in an alarm condition. The potential health risk involved in this state can be the discontinuance of therapy.
Root causeOther
ActionAlsius Corporation is notifying customers of an issue requiring service of their CoolGard 3000 by letter shipped via FEDEX and DHL. Verification of receipt of the notification letter to level A will be undertaken by using tracked delivery requiring signature upon receipt.
Quantity239
DistributionNationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malaysia, Italy, Ireland, Greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas, and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K014241COOLGARD 3000, COOLINE, START-UP, MODELS 3000, CL-2085B, CL-2295A, CG-500D
Alsius Corp., Aug 1, 2003
K030421ALSIUS COOLGARD 3000 & ALSIUS CATHETER SYSTEM
Alsius Corp., Oct 23, 2003