Recall Z-0347-06 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Renal Div

System 1000 family of Hemodialysis Instruments, including the System 1000, TINA, AURORA and ARENA; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

DateJan 6, 2006 (initiated Nov 18, 2005)
StatusTerminated
ReasonPotential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.
Root causeOther
ActionBaxter sent Urgent Device Correction letters dated 11/18/05 to all System 1000, TINA, Aurora and Arena Hemodilaysis Instrument customers via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the potential exists for fluid or air to be passed through the venouus line clamp that could result in an air emboli condition. This condition can occur when the tubing is not centered on the clamp and extends beyond the edge of the clamp''s pinch zone. Baxter will make arrangements to upgrade the instruments with either a linea
Quantity29,173 machines
DistributionNationwide and internationally to Canada, Korea, Hong Kong, Shanghai, Jamaica, Philippines, United Kingdom, Denmark, India, Belgium, Mexico, Italy, Austria, Sweden, Ecuador, Argentina, Brazil, Chile, Colombia, Peru, Turkey and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013562AURORA SYSTEM 1000 SERIES HOME SELF-CARE SINGLE PATIENT DELIVERY SYSTEM; MODELS- SYS 1000, L3 WITH N100 ARM, 4B SOFTWARE
Baxter Healthcare Corp, Dec 12, 2002
K030099BAXTER ARENA HEMODIALYSIS DELIVERY SYSTEM, MODEL 1571278000
Baxter Healthcare Corp, Jun 17, 2003
K910215DRAKE WILLOCK SYSTEM 1000 DIALYSATE DELIVERY SYST
Cd Medical, Inc., Apr 9, 1991