ClearanceBrief
Recall Z-0388-06 ยท ISH Component, External, Limb, Ankle/Foot

Ossur

Elation, Class I Medical Device Prosthetic product, Component, External, Limb, Ankle/Foot Parts Numbers: ELP0023L/R-ELP0026L/R, ELP0023RBR, ELP0024RBR &, ELP0028RBR.

DateJan 12, 2006 (initiated Oct 21, 2005)
StatusTerminated
ReasonAfter one year devices assembled with a new type of O-rings could exhibit enough wear to allow a gradual oil leak which would increase play in the ankle sub-assembly of the foot.
Root causeOther
ActionTelephone communication followed by notification letter mailed to end users (patients) and prosthetics workshops. A pre-paid package to return faulty components to Ossur included with mailed letters. New replacement feet will be shipped to customers.
Quantity48 (23 in US)
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall