Recall Z-0266-06 ยท DXY Implantable Pacemaker Pulse-Generator

ela Medical Llc

Symphony DR model 2550 and ELA Rhapsody DR model 2510. A dual chamber rate-responsive pacemaker that uses both minute ventilation and accelerometer sensors. Sterilization with Ethylene Oxide Gas. Sorin Group, ELA medical, 98, rue Maurice Amoux, 92120 Montrouge - France. Made in France.

DateDec 2, 2005 (initiated Oct 25, 2005)
StatusTerminated
ReasonThrough post-market surveillance, ELA Medical has observed that a no-output condition could occur in a limited number of Symphony or rhapsody pacemakers. The no-output condition could occur due to metal migration caused by a specific manufacturing process.
Root causeOther
ActionA Dear Doctor letter, dated 10/25/05, was sent to all affected doctors. The letter describes the failures and mentions that pacemaker depend patients implanted with devices manufactured in the first group could require prophylactic replacement.
Quantity17
DistributionDevices distributed with the US to AZ, CA, FL, IL, LA, MA, ME, MO, MS, SC,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.