ClearanceBrief
Recall Z-0179-06 ยท HAW Neurological Stereotaxic Instrument

FHC, Inc.

FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.

DateNov 23, 2005 (initiated Oct 31, 2005)
StatusTerminated
ReasonPower Assist microTargeting System may advance beyond intended target
Root causeOther
ActionFHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movem
Quantity70 units
DistributionNationwide Foreign: South Africa, Taiwan, Hong Kong, China, France
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011992FHC MICROTARGETING DRIVE SYSTEM
FHC, Inc., Aug 14, 2001