Recall Z-0274-06 ยท KPR System, X-Ray, Stationary

Eastman Kodak Co

KODAK DirectView DR 7100 System, Catalog # 133-6049. This is a digital radiology system. The major components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.

DateDec 8, 2005 (initiated Oct 20, 2005)
StatusTerminated
ReasonPotential weld defect in the Overhead Tube Support assembly which connects the tube assembly to the telescoping arm. This could cause the tube assembly to fall, injuring anyone near the unit.
Root causeOther
ActionLetters dated 10/20/2005 were hand carried by Kodak Field Service Reps. Kodak Field Engineers inspected units and found the two defective units (both in China).
Quantity3
DistributionDOMESTIC: 12 units shipped to 12 medical facilities in FL, KY, MI, NY, OH, PA, and SC. FOREIGN: 12 units shipped to 7 hospitals in China, 1 warehouse in China; 1in the possession of Chinese Customs (government); and 1 is in a warehouse in Russia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001341DR5000 DEDICATED CHEST X-RAY SYSTEM
Analogic Corp., Jun 2, 2000