Recall Z-0138-06 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735

DateNov 15, 2005 (initiated Oct 5, 2005)
StatusTerminated
ReasonIf the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an incorrect location and presents the possibility of an inconclusive diagnosis. calculation of offset values for the breast biopsy aiming device may be incorrect
Root causeOther
ActionThe recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions MR034/05/S. The letter re-emphasizes the importance of printing the offset values on the worksheet prior to moving the patient table and in accordance with the User's Manual.
Quantity16 units
DistributionThe product was shipped to medical facilities in CA, FL, IL, MD, MN, NM, OH, OR, RI, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010773BREAST BIOPSY DEVICE/MAGNETOM SYSTEMS
Siemens Medical Solutions USA, Inc., May 30, 2001