ClearanceBrief
Recall Z-0550-06 ยท MEA Pump, Infusion, Pca

Baxter Healthcare Corp.

AP II Pain Management System infusion pump, product codes 2L3105, 2L3105K, 2L3105R, 2L3105T and 2L3105W; Baxter Healthcare Corporation, Deerfield, IL 60015

DateApr 26, 2006 (initiated Oct 27, 2005)
StatusTerminated
ReasonWhen infusing at flow rates below 0.5 mL/hr, the upstream occlusion alarm of the pumps may not detect the occlusion.
Root causeOther
ActionBaxter sent Urgent Device Correction letters dated 10/27/05 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for air to migrate through the silicone-pumping segment of the sets, which may result in air reaching the patient. Baxter recommended that the users of the I-Pump to enable the upstream occlusion sensor of the pump or to use sets containing air-elimination filters if the occlusion sensor is disabled or otherwise not activated. Baxter also recommended that for delivery routes other than epidural and subcutaneous at flow r
Quantity20,470 pumps
DistributionNationwide, including Puerto Rico, and internationally to Canada, Chile and China
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K926385BARD APII PUMP
C.R. Bard, Inc., Apr 18, 1994