Recall Z-0194-05 ยท DXY Implantable Pacemaker Pulse-Generator
Guidant Corporation
Guidant INSIGNIA I AVT family of pacemakers includes the following: SSI (model 482), VDD (model 882), DDD (model 982), SR (model 1192 and DR (model 1292). Intermedics NEXUS I AVT family of pacemakers includes the following: SR (model 1392). The INSIGNIA I AVT pacemakers are multiprogrammable pacemakers from Guidant. The NEXUS I AVT pacemakers are multiprogrammable pacemakers from Intermedics. The
| Date | Nov 23, 2005 (initiated Sep 22, 2005) |
| Status | Terminated |
| Reason | Another failure mode, which occurs at a low rate, has been identified with the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include; intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation. |
| Root cause | Other |
| Action | Important Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05. UPDATE: Guidant is issuing an Advisory Update to physicians in December 2005. This update provides information relative to the failure mode including the identified cause; FDA's classification; Guidant representatives will retrieve and replace any hospital inventory with product free from "Mode 2" peri-implant failure; and recommended actions |
| Quantity | 14,404 |
| Distribution | Nationwide and Arab Emerites, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.