ClearanceBrief
Recall Z-0193-06 ยท DXY Implantable Pacemaker Pulse-Generator

Guidant Corporation

Guidant INSIGNIA I Plus family of pacemakers includes the following: SR (model 1194); and DR (models1297, 1298). Intermedics NEXUS I Plus family of pacemakers includes the following: SR (model 1394) and DR (model 1467, 1468). The INSIGNIA I Plus pacemakers are multiprogrammable pacemakers from Guidant. The NEXUS I Plus pacemakers are multiprogrammable pacemakers from Intermedics. The family consis

DateNov 23, 2005 (initiated Sep 22, 2005)
StatusTerminated
ReasonAnother failure mode, which occurs at a low rate, has been identified with the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include; intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.
Root causeOther
ActionImportant Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05. UPDATE: Guidant is issuing an Advisory Update to physicians in December 2005. This update provides information relative to the failure mode including the identified cause; FDA's classification; Guidant representatives will retrieve and replace any hospital inventory with product free from "Mode 2" peri-implant failure; and recommended actions
Quantity174,848
DistributionNationwide and Arab Emerites, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.