Recall Z-0191-06 ยท DXY Implantable Pacemaker Pulse-Generator

Guidant Corporation

Guidant INSIGNIA I Entra family of pacemakers includes the following: SSI (model numbers 0484 and 0485); DDD (models 0985, 0986); SR (models 1195, 1198); and DR (models1294, 1295, 1296). Intermedics NEXUS Entra family of pacemakers includes the following: SSI (model 1326); DDD (model 1426); SR (model 1398) and DR (model 1494). The INSIGNIA I Entra pacemakers are multiprogrammable pacemakers from G

DateNov 23, 2005 (initiated Sep 22, 2005)
StatusTerminated
ReasonAnother failure mode, which occurs at a low rate, has been identified with the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include; intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.
Root causeOther
ActionImportant Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05. UPDATE: Guidant is issuing an Advisory Update to physicians in December 2005. This update provides information relative to the failure mode including the identified cause; FDA's classification; Guidant representatives will retrieve and replace any hospital inventory with product free from "Mode 2" peri-implant failure; and recommended actions
Quantity93,896
DistributionNationwide and Arab Emerites, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea
FDA recordOpen on fda.gov