ClearanceBrief
Recall Z-0093-06 ยท HQL Intraocular Lens

Advanced Medical Optics, Inc.

Tecnis Monofocal Silicone Posterior Chamber IOLs (Models: Z9000 and Z9001)

DateOct 28, 2005 (initiated Oct 5, 2005)
StatusTerminated
ReasonA small opening (channel or tunnel) in the seal of the outer Tyvek pouch which contains the lens case, in which the sealed intraocular lens is packaged may not be adequately sealed, therefore the potentially compromised outer Tyvek pouch may not be sterile and may potentially contaminate an otherwise sterile field during surgery.
Root causeOther
ActionA product notification letter was mailed to all US domestic customers via overnight carrier (Federal Express) beginning on October 5, 2005. A global notification letter will be translated into the local languages by the international AMO affiliates and mailed to international customers as soon as a listing of all customers is tabulated.
Quantity107 (39 in US)
DistributionNationwide, Austria, Belgium, Switzerland, Germany, Denmark, France, UK, Guadelope, Hong Kong, Croatia, India, Italy, Netherlands, and Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.