Recall Z-0137-06 ยท KHO Fluorometer, For Clinical Use

Ortho-Clinical Diagnostics

Vitros ECi Immunodiagnostic System, sold as Cat. #863 3893; Cat. #192 2814 (Vitros ECiQ), and Cat. #680 1059 (recertified Vitros ECi) when using Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs. NOTE: These 3 catalog numbers of Vitros ECi represent the original unit, an updated unit with cosmetic changes only, and refurbished units.

DateNov 15, 2005 (initiated Sep 29, 2005)
StatusTerminated
ReasonWhen using Vitro ECi Immunodiagnostic System with Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs, results of the 1st of two required re-test replicates may be negtively biased.
Root causeOther
ActionLetters dated 9/29/2005 issued to all domestic customers of the VITROS ECi/ECiQ systems and to Ortho/J&J foreign affiliates. Customers instructed to immediately discontinue using reflex processing to the VITROS Anti-HBc IgM and VITROS Anti-HAV reagent packs.
Quantity3,507 Vitros ECi units distributed
DistributionNationwide and worldwide distribution of the VITROS ECi Immunodiagnostic systems.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962919VITROS IMMUNODIAGNOSTIC SYSTEM
Johnson & Johnson Clinical Diagnostics, Inc., Oct 18, 1996