Recall Z-0154-06 ยท DXT Injector And Syringe, Angiographic

Medrad Inc

Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX

DateNov 16, 2005 (initiated Oct 10, 2005)
StatusTerminated
Reasonthe heat maintainer, an accessory, may overheat and deform the disposable syringe
Root causeOther
ActionThe recalling firm issued an Urgent Medical Device Field Correction notice dated 10/12/05 via certified mail. The notice provided instructions to apply warning labels and disconnect the heat maintainer until a field upgrade kit is available. In case where the heat maintainer is a necessity for specific procedures, the customer is being instructed to observe certain precautions.
Quantity4362 units
DistributionThe products were shipped to medical facilities nationwide. The products were also shipped to Canada, Isreal, Italy, Denmark, Netherlands, Norway, Great Britain, France, Poland, Germany, Spain, Finland, Belgium, Sweden, Russia, Austraila, Cuba, Mexico, Peru, Hong Kong, Korea, Mylasia, Thialand, Taiw
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023183MEDRAD STELLANT CT INJECTOR SYSTEM
Medrad, Inc., Jan 10, 2003
K033881MEDRAD STELLANT CT INJECTOR SYSTEM WITH IMAGING SYSTEM INTERFACE MODULE
Medrad, Inc., Mar 2, 2004