Recall Z-0105-06 ยท KPS System, Tomography, Computed, Emission

General Electric Med Systems LLC

Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System

DateOct 28, 2005 (initiated Sep 27, 2005)
StatusTerminated
ReasonCracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.
Root causeOther
ActionGE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.
Quantity140 (80 US and 60 OUS)
DistributionWithin the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K942837DST-XL
Sopha Medical Systems, Inc., Jan 5, 1995