Recall Z-0147-06 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

St Jude Medical

Photo Micro VR/DR Models V-194

DateNov 16, 2005 (initiated Oct 6, 2005)
StatusTerminated
ReasonVendor-supplied memory chip in a limited number of St. Jude Medical ICDs can trigger a temporary loss of pacing function and permanent los of defibrillation support.
Root causeOther
ActionNotification material sent to physicians by FedEx. Acknowledgement of receipt of the notification material will be monitored with this information being fed back to St. Jude Medical CRMD in Sylmar, California. In addition, St Jude Medical issued a press release on October 7, 2005.
Quantity7696
DistributionNationwide and Western Europe: Austria Belgium Denmark Finland France Germany Italy Netherlands Poland Portugal Spain Sweden UK Switzerland Greece Malta Eastern Europe: Czech Republic Lithuania Latvia Romania Bulgaria Croatia Slovenia Slovakia Russia Serbia Middle East!