ClearanceBrief
Recall Z-0114-06 ยท MVC System, Test, Her-2/Neu, Ihc

Dakocytomation California Inc

K 520711-2, Hercep Test, 35Test, HER2, Protein

DateNov 1, 2005 (initiated Sep 29, 2005)
StatusTerminated
ReasonFormulation error in the Visualization Reagent (vial 3) component contained in Hercep Test kits. The error was in the concentration of Polyethylene Glycol contained in the formulation.
Root causeOther
ActionNotification letter dated Sept 29,2005 was sent by certified mail to consignees instructing disposal of product.
Quantity420
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.