Recall Z-0892-06 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Guidant Corporation

VITALITY DS (model T125) and VITALITY 2 (models T165, 175) ICD (implantable cardiac defibrillator)

DateMay 23, 2006 (initiated Sep 30, 2005)
StatusTerminated
ReasonDevices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.
Root causeOther
ActionGuidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patie
Quantity40
DistributionNationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.