Recall Z-0198-06 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Toshiba American Med Sys Inc

High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

DateNov 24, 2005 (initiated Sep 21, 2005)
StatusTerminated
ReasonIt was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.
Root causeOther
ActionConsignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.