Recall Z-0009-06 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc

Synchron LX Systems Version 4.5 Operating Software

DateOct 6, 2005 (initiated Sep 6, 2005)
StatusTerminated
ReasonSynchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.
Root causeSoftware design
ActionCustomers were notified by letter dated September 6, 2005 with work around instruction provided in the letter.
Quantity868
DistributionNationwide and Canada