Recall Z-0339-06 ยท GWM Device, Monitoring, Intracranial Pressure

Integra Neuro Sciences

NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit

DateJan 6, 2006 (initiated Sep 8, 2005)
StatusTerminated
ReasonCertain Model NS-P probes contained in the NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit may allow an Intracranial Pressure (ICP) drift of up to 5 mmHg during the first 24 hours of monitoring ICP. This is outside the products labeling specification of 2 mmHg maximum ICP drift within the first 24 hours of monitoring.
Root causeOther
ActionNotify Customers and Sales Representatives who received the product by letter sent via Federal Express. Customers will be instructed that an Integra Sales Representative will contact them to arrange for return of affected product.
Quantity79
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050720NEUROSENSOR CEREBRAL BLOOD FLOW AND INTRACRANIAL PRESSURE MONITORING KIT/SYSTEM, MODELS NS-P, NS-100
Integra Neurosciences, May 11, 2005