Recall Z-0111-06 ยท KRA Catheter, Continuous Flush

Ekos Corp

Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System c

DateNov 1, 2005 (initiated Nov 22, 2004)
StatusTerminated
ReasonDuring two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In one case, the USC fractured within the DDC when the user attempted to withdraw the USC after resistance was encountered
Root causeOther
ActionOn 11/22/04 the firm issued letters to those customers that had product. The letters advised of the problem and provided amended use instructions.The firm continued to ship product and on 3/10/05 the firm visited consignees, replacing the catheters.
Quantity134 kits
DistributionKits were distributed to 6 hospitals in the US, 1 each located in AZ, OH, WA, FL, IL, and CA. The firm also distributed to a customer in Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042456Lysus Infusion System
Ekos Corp., Sep 23, 2004