Recall Z-0110-06 ยท KRA Catheter, Continuous Flush

Ekos Corp

The Lysus Infusion System - Catheter Interface Cable. The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System cons

DateNov 1, 2005 (initiated Sep 21, 2004)
StatusTerminated
ReasonThe 100V board in the catheter interface cable pod failed during use. The PT-3 control unit would not recognize the catheter interface cable.
Root causeOther
ActionOn 9/21/04 the firm began visiting customers and replacing the catheter interface cables.
Quantity3
DistributionUnits were shipped to hospitals, one each in WA, AZ, and OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042456Lysus Infusion System
Ekos Corp., Sep 23, 2004