ClearanceBrief
Recall Z-0223-06 ยท DPS Electrocardiograph

Smiths Medical PM, Inc.

Advisor Vital Signs Monitor (model 9200), catalog number 920654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main boards (P/N: 80002B1)3-lead 60 Hz ECG, standarn SPO2, Invasive pressure/temperature option, printer and battery.

DateNov 30, 2005 (initiated Sep 1, 2005)
StatusTerminated
ReasonPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
Root causeOther
ActionDistirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004. Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB). The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.
Quantity1
DistributionAZ, CA, KS, KY, NY, TN, TX OUS to countries such as: Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010770BCI ADVISOR VITAL SIGNS MONITOR (9200)
Bci, Inc., Jun 12, 2001
K011177BCI ADVISOR VITAL SIGNS MONITOR (9200)
Bci, Inc., May 23, 2001
K030407ADVISOR, MODEL 9200
Bci, Inc., Aug 19, 2003
K982279VITAL SIGNS MONITOR; ADVISOR MODEL # BCI 9200
Bci Intl., Inc., Sep 25, 1998