ClearanceBrief
Recall Z-0023-06 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Zimmer Inc.

Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface, small, sterile, for cemented use only in the USA, Cat. no. 00430604000.

DateOct 7, 2005 (initiated Aug 23, 2005)
StatusTerminated
ReasonThe implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.
Root causeOther
ActionConsignees were notified via letter dated August 23, 2005 or by email.
DistributionNationwide, Australia, Canada, England, Germany and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022377THE TRABECULAR METAL GLENOID-BIGLIANI/FLATOW THE COMPLETE SHOULDER SOLUTION, MODEL 4306
Implex Corp., Dec 10, 2002
K031449MODIFICATION TO THE TRABECULAR METAL GLENOID - BIGLIANI/FLATOW THE COMPLETE SHOULDER SOLUTION
Implex Corp., Sep 23, 2003