Recall Z-0387-06 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

General Electric Med Systems LLC

LOCIQ Book System models: 2399921, 2399923, 2349933, 2349934 and 2349935

DateJan 11, 2006 (initiated Sep 2, 2004)
StatusTerminated
ReasonThe Estimated Fetal Weight (EFW) result will be incorrect if the operator fails to perform the 'end exam' or new patient registration step before starting an EFW measurement procedure on the next patient.
Root causeOther
ActionDevices will have software upgrades done by GE Field Service Technicians. The upgrade will include a dialog box which will display on the screen when prior data exists and a new patient is being registered.
Quantity2698
Distributionnationwide and worldwide