Recall Z-0557-06 ยท LJY Enzyme Linked Immunoabsorbent Assay, Mumps Virus

Diasorin Inc.

The ETI-Mumps IgG Enzyme Immunoassay Test Kit is sold with assay software version 1.0. that is designed to work exclusively with the ETI-Max 3000 automated platform (version 1.51.1).

DateFeb 23, 2006 (initiated Aug 8, 2005)
StatusTerminated
ReasonPotential for false negative results. A software anomaly may occur when performing the ETI-Mumps IgG assay on the ETI-Max 3000 automated platform. The error occurs when running a combined plate on the ETI-Max 3000, with the ETI-Mumps IgG assay as the second assay on a combined plate with ETI-Measles IgG and/or ETI-VZV IgG.
Root causeOther
ActionConsignees were contacted by telephone and then the firm sent a fax titled 'Diasorin Customer Notification' dated 8/8/05 that documented the information as provided by telephone.
Quantity22
DistributionNationwide. AK, CA, CT, DC, KS, MA, NY, NJ, PA, SD, TN, TX.